Title of article
A single photon emission computer tomograph for breast cancer imaging
Author/Authors
Vecchio، نويسنده , , S. and Belcari، نويسنده , , N. and Bennati، نويسنده , , P. and Camarda، نويسنده , , M. and Campanini، نويسنده , , R. and Cinti، نويسنده , , M.N. and Del Guerra، نويسنده , , A. and Iampieri، نويسنده , , E. and Lanconelli، نويسنده , , N. and Pani، نويسنده , , R. and Spontoni، نويسنده , , L.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2007
Pages
4
From page
84
To page
87
Abstract
We have developed a tomograph for single photon emission imaging (SPECT) of the breast for the detection of small size tumors. The SPECT is mounted on a ring that is rotating around the breast with the patient in prone position. The breast will be imaged by two opposing detector heads of approximately 5 × 15 cm 2 each, with a field of view about 13 cm wide. Each head is made up of one pixilated NaI crystal matrix coupled to three Hamamatsu H8500 PMTs. A “general purpose” lead collimator is positioned in front of the crystal. Detailed simulations have been made for the optimization and the evaluation of the detector performances. Monte Carlo results indicate that tumors of 8 mm diameter are detectable with a tumor/background uptake ratio of 5:1. The experimental characterization of the detector head is presented. The rotating ring is now being assembled.
Keywords
SPECT , MA-PMT , breast cancer
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Serial Year
2007
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Record number
2208051
Link To Document