• Title of article

    Empirical dead-time corrections for energy-resolving detectors at synchrotron sources

  • Author/Authors

    Walko، نويسنده , , D.A. and Arms، نويسنده , , D.A. and Miceli، نويسنده , , A. and Kastengren، نويسنده , , A.L.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2011
  • Pages
    3
  • From page
    81
  • To page
    83
  • Abstract
    We examine the high count-rate performance of an energy-resolving detector in the three operating modes of the Advanced Photon Source (APS). Specifically, we present the optimal dead-time corrections for the SII Vortex silicon drift diode (SDD) detector using a digital pulse processor, highlighting the differences in operation between the 24-bunch, 324-bunch, and hybrid singlet modes of the APS. We analyze the input count rate (ICR), output count rate (OCR), and several regions of interest (ROIs). We find that the correct formula for dead-time correction can extend the use of the detector to significantly higher count rates.
  • Keywords
    Fluorescence detector , efficiency , Dead-time , Synchrotron
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2011
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number