• Title of article

    Improved performance in synthetic diamond neutron detectors: Application to boron neutron capture therapy

  • Author/Authors

    Almaviva، نويسنده , , Laura S. and Marinelli، نويسنده , , Marco and Milani، نويسنده , , E. and Prestopino، نويسنده , , G. and Tucciarone، نويسنده , , A. and Verona، نويسنده , , C. and Verona-Rinati، نويسنده , , G. and Angelone، نويسنده , , M. and Pillon، نويسنده , , M.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2010
  • Pages
    3
  • From page
    580
  • To page
    582
  • Abstract
    An improved thermal and fast neutrons detector is obtained, modifying a recently proposed multilayered homoepitaxial Chemical Vapor Deposition (CVD) diamond detector (M. Marinelli, et al., Appl. Phys. Lett. 89 (2006) 143509), where a 6LiF layer deposited on the sensing layer was used to convert thermal neutrons into charged particles. By sandwiching this layer between two CVD diamond detectors connected in parallel, a better signal-to-background separation is achieved. This allows to use 10B as converting element, so to realize a detector suitable for Boron Neutron Capture Therapy dosimetry. Also, the doubled detector volume enhances the sensitivity to fast neutrons.
  • Keywords
    diamond , dosimetry , detector , BNCT
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2010
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number