Measurement and detailed simulation of the Modulation Transfer Function of the Medipix2
Author/Authors
Bartl، نويسنده , , Peter and Korn، نويسنده , , Alexander and Michel، نويسنده , , Thilo and Anton، نويسنده , , Gisela، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2008
Pages
4
From page
314
To page
317
Abstract
The Modulation Transfer Function (MTF) is an important quality characteristic for the description of the spatial resolution of imaging systems. For X-ray detectors, like the Medipix2, it depends for instance on the pixel size and the incident spectrum. The detection of single photons in the Medipix2 is additionally influenced by the lower energy threshold. The dependency of the MTF on this threshold and on the incident spectrum has been evaluated by simulations and measurements. By choosing an adequate threshold, the so-called low frequency drop can be suppressed or the spatial resolution in general is influenced.
Keywords
Edge-method , Low frequency drop , Medipix2 , Pixel detector , MTF