• Title of article

    Experimental comparison of different dead-time correction techniques in single-channel counting experiments

  • Author/Authors

    Michotte، نويسنده , , C. and Nonis، نويسنده , , M.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2009
  • Pages
    6
  • From page
    163
  • To page
    168
  • Abstract
    For single-channel counting, the dead-time correction method using transmission formulae was compared experimentally with the live-time technique. Agreement is within 4×10−4 for count rates up to 104 s−1 when the formulae for dead times in series are used and the dead-time value related to the analogue pulse width is adjusted for the BIPM modules. At higher count rates, a bias of 2×10−3 is observed. Two different live-time modules were compared and it was demonstrated that the live-time correction is also sensitive to the analogue pulse width and to the duration or rise time of the clock pulses. When these effects are taken into account, the two modules differ by only 5×10−4.
  • Keywords
    Dead times in series , Dead-time transmission formulae , Analogue pulse width , Live-time correction , (Non-)extendable dead time
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2009
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number