Title of article
Fabrication and performance tests of a segmented p-type HPGe detector
Author/Authors
King III، نويسنده , , George S. and Avignone III، نويسنده , , Frank T. and Cox، نويسنده , , Christopher E. and Hossbach، نويسنده , , Todd W. and Jennings، نويسنده , , Wayne and Reeves، نويسنده , , James H.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2008
Pages
6
From page
599
To page
604
Abstract
A p-type semi-coaxial HPGe detector has been segmented by cutting, with a diamond saw, and etching four circumferential grooves through the Li-diffused dead layer. The degree of segmentation was tested using a well-collimated low-energy gamma-ray source. An analysis cut that rejected events depositing energy in more than one segment was applied to an energy interval of 2038±5 keV, the region of interest (Qββ) for 76Ge 0νββ decay experiments. This segmentation cut resulted in a reduction of the Compton continuum of 59%.
Keywords
Segmented germanium detectors
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Serial Year
2008
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Record number
2210384
Link To Document