• Title of article

    After-pulsing and cross-talk in multi-pixel photon counters

  • Author/Authors

    Du، نويسنده , , Y. and Retière، نويسنده , , F.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2008
  • Pages
    6
  • From page
    396
  • To page
    401
  • Abstract
    Multi-pixel photon counters (MPPC) are pixelated Geiger-mode photon manufactured by Hamamatsu photonics. They will be used for reading out all the scintillator elements within the near detector of the T2K experiment. Their performances photo-detection efficiency, dark noise and gain, and their insensitivity to magnetic field fulfill the ND280 requirements. On the other hand, two known issues, cross-talk and after-pulsing may adversely impact the detector response. In this paper, cross-talk and after-pulsing are precisely measured by recording waveforms and identifying all the avalanche pulses. At an over-voltage of 2 V, an avalanche generates on an average 0.5 additional avalanches due to after-pulsing, whereas it generates 0.13 at 1 V over-voltage. After-pulses follow two independent time constants of 15 and 82 ns. The cross-talk probability is a factor of 3 smaller than the after-puling probability. Simulating the MPPC response, we find that after-pulsing and cross-talk do not degrade the detector performance significantly within the expected operating conditions of the T2K near detector.
  • Keywords
    scintillator , Photo-detector , GPD , MPPC , Avalanche photodiode
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2008
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number

    2210455