Characterization and selection of CZT detector modules for HEX experiment onboard Chandrayaan-1
Author/Authors
Vadawale، نويسنده , , S.V. and Purohit، نويسنده , , Sathyanarayanan S. and Shanmugam، نويسنده , , M. and Acharya، نويسنده , , Y.B. and Goswami، نويسنده , , J.N. and Sudhakar، نويسنده , , M. and Sreekumar، نويسنده , , P.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2009
Pages
11
From page
485
To page
495
Abstract
We present the results of characterization of a large sample of Cadmium Zinc Telluride (CZT) detector modules planned to be used for the HEX (High Energy X-ray spectrometer) experiment onboard Indiaʹs first mission to the Moon, Chandrayaan-1. We procured forty modules from Orbotech Medical Solutions Ltd. and carried out a detailed characterization of each module at various temperatures and selected final nine detector modules for the flight model of HEX. Here we present the results of the characterization of all modules and the selection procedure for the HEX flight detector modules. These modules show 5–6% energy resolution (at 122 keV, for best 90% of pixels) at room temperature which is improved to ∼4% when these modules are cooled to sub-0 °C temperature. The gain and energy resolution were stable during the long duration tests.
Keywords
CZT , CdZnTe , Solid-state X-ray detector , Pixel array