Title of article
Development of an ultra-high-speed scanning neutron tomography system for high-quality and four-dimensional visualizations
Author/Authors
Kureta، نويسنده , , M. and Iikura، نويسنده , , H.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2009
Pages
4
From page
81
To page
84
Abstract
A new neutron tomography imaging system was developed in order to realize high-quality three-dimensional (3D) and four-dimensional (4D) visualizations by fusing the high-frame-rate neutron radiography and computed tomography (CT) techniques. The fundamental idea is that the object is revolved with high rotating speed and the neutron radiography images are recorded with a high-speed video camera and an image intensifier, and then the consecutive images are processed by a 3D CT technique. The 4D dynamic images of the sand flow in the sandglasses could be visualized clearly. This new technique has also an advantage in that it can reduce radio-activation of the object materials remarkably.
Keywords
Neutron , Neutron tomography , Measurement , Four-dimensional , CT , High-speed imaging , Visualization , Neutron radiography
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Serial Year
2009
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Record number
2211338
Link To Document