    Optimizing the design of a moderator-based neutron detector for a flat response curve in the 2–14 MeV energy range

  • Author/Authors

    Hu، نويسنده , , Q.Y. and Li، نويسنده , , B.J. and Zhang، نويسنده , , D. and Guo، نويسنده , , H.S. and Yang، نويسنده , , G.Z. and Si، نويسنده , , F.N. and Liu، نويسنده , , J.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2009
  • Pages
    4
  • From page
    213
  • To page
    216
  • Abstract
    The efficiency of a neutron detector with boron trifluoride proportional tubes embedded in a polyethylene moderator was simulated with a Monte Carlo program. A moderator structure where the detector had uniform sensitivity for neutrons from 2 to 14 MeV was determined by simulation. A counter was built based on the simulation results. The counterʹs efficiencies were calibrated with an Am–Be source and an accelerator that served as a D–D and D–T neutron source. Experimental neutron efficiencies of these sources are approximately uniform. The simulated model was validated by the consistent results between the calculated and experimental data.
  • Keywords
    Energy response , Moderator , Monte Carlo simulation
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2009