Title of article
Experimental estimation of the accuracy of modern scalar quantum magnetometers in measurements of the Earthʹs magnetic field
Author/Authors
Shifrin، نويسنده , , V.Ya. and Khorev، نويسنده , , V.N. and Kalabin، نويسنده , , V.N. and Park، نويسنده , , P.G.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2008
Pages
6
From page
147
To page
152
Abstract
The results of calibration of proton, cesium, potassium and helium–cesium scalar magnetometers in the range 20–100 μT are discussed. The calibration made it possible to estimate errors and to apply corrections to reduce the uncertainty of measurements.
mental data have been obtained using a standard calibration system designed in the Russian Mendeleyev Metrology Institute (VNIIM). The systematic uncertainty in the reproduction of the magnetic flux density is estimated to be 1.8 × 10−7, the random standard deviation being 0.001 nT.
perimental data can be used for the proper selection of measuring instruments and prediction of their uncertainty when studying the Earthʹs magnetic field.
Keywords
Measurement , uncertainty , magnetic resonance , Calibration , Magnetometer
Journal title
PHYSICS OF THE EARTH AND PLANETARY INTERIORS
Serial Year
2008
Journal title
PHYSICS OF THE EARTH AND PLANETARY INTERIORS
Record number
2304787