• Title of article

    Introducing a Dedicated Cardiac SPECT Imaging System: ProSPECT

  • Author/Authors

    Teimourian Fard, Behnoosh Research Center for Molecular and Cellular Imaging - Tehran University of Medical Sciences , Amirmozaffari Sabet, Kiarash School of Electrical Engineering - Sharif University of Technology , Ay, Mohmmadreza Department of Medical Physics and Biomedical Engineering - Tehran University of Medical Sciences

  • Pages
    5
  • From page
    156
  • To page
    160
  • Abstract
    Purpose: Cardiac PSECT imaging is widely used for many clinical practices such as the diagnostic of coronary artery diseases (CAD) and myocardial perfusion imaging (MPI). In this work, we introduced the ProSPECT system, a dedicated cardiac SPECT system with open-gantry design. Materials and Methods: In we assessed the performance of the system based on NEMA-NU1-2007 standards. The ProSPECT system was characterized by measurement of planar, tomographic and clinical parameters.Results: Planar measurements showed that 7.6 mm and 8.9% of spatial resolution and energy resolution at 140 keV, respectively. The tomographic resolution at 250 mm radius of rotation was 12.3 mm. Also, the maximum LOR and COR error for the tomographic imaging is obtained 1.1 mm and 1.8 mm respectively. The clinical images that obtained using the system confirmed by expert nuclear medicine physician. Conclusion: It concluded that the ProSPECT imaging is qualified for cardiac SPECT imaging.
  • Keywords
    Cardiac SPECT , Nuclear Medicine , NEMA , Calibration
  • Journal title
    Frontiers in Biomedical Technologies
  • Serial Year
    2019
  • Record number

    2500502