Title of article
Optimized Blood Volume Monitoring during Hemodialysis Procedure based on Ultrasonic Speed Measurement
Author/Authors
Nafisi, V. R Biomedical Engineering Group - Iranian Research Organization for Science & Technology, Tehran, Iran , Eghbal, M Biomedical Engineering Group - Iranian Research Organization for Science & Technology, Tehran, Iran
Pages
8
From page
373
To page
380
Abstract
Fast reduction of blood water volume due to ultra-filtration via hemodialysis machine
and critical threshold of blood water volume can result in clinical complications such
as hypotension. When blood water volume reduces during hemodialysis process, the
concentration of blood forming elements will increase. One of the methods for the
monitoring of the concentration changes is ultrasonic speed measurement.
In this paper, we optimize the ultrasonic speed measurement method in such a way
that it would be compatible with conventional air detector in a hemodialysis machine.
Air bubble detecting and the relative blood water volume estimating can be achieved
by adding the “sing-around” ultrasonic speed measurement method to the current hemodialysis
machine; air bubble detection is achieved by monitoring the decline in
the ultrasonic wave amplitude (because of air bubbles) and the protein concentration
measurement is achieved by measuring the transit time of the wave. Therefore, the optimized
method can be applied to the air detection device of a hemodialysis machine.
Keywords
Air Detector , Ultrasonic Speed Measurement , Blood Volume Monitoring , Hemodialysis
Journal title
Journal of Biomedical Physics and Engineering
Serial Year
2019
Record number
2500646
Link To Document