Title of article
Precision of Low-Dose Response of LiF:Mg, Ti Dosimeters Exposed to 80 kVp X-Rays
Author/Authors
Bauk, Sabar Universiti Sains Malaysia - School of Distance Education - Physical Sciences Programme, Malaysia , Alam, Md. Shah Universiti Sains Malaysia - School of Distance Education - Physics Programme, Malaysia , Alam, Md. Shah Shahjalal University of Science and Technology - Department of Physics, Bangladesh , Alzoubi, Ahmad Saleem Universiti Sains Malaysia - Advanced Medical and Dental Institute, Malaysia
From page
125
To page
130
Abstract
The low-dose response of a LiF:Mg, Ti thermoluminescent dosimeter (TLD-100) was studied by exposing the TLDs to 80 kVp X-rays to different doses. There was no evidence of non-linearity in the dose range studied. The standard deviation was less than 10% at 0.3 mGy, the lowest thermoluminescent (TL) dose before the uncertainty in the TL reading became very large, which reached approximately 60%.
Keywords
TL , dose response , diagnostic x , ray , low dose
Journal title
Journal of Physical Science
Journal title
Journal of Physical Science
Record number
2581538
Link To Document