Title of article
Reliability Assessments of Infant Incubator and the Analyzer
Author/Authors
ÖZDEMIRCI, Emre Çankırı Karatekin University - Technical and Business College, Turkey , ÖZARSLAN YATAK, Meral Gazi Üniversitesi - Faculty of Technical Education, Turkey , DURAN, Fecir Gazi Üniversitesi - Faculty of Technology, Turkey , CANAL, Mehmet Rahmi Gazi Üniversitesi - Faculty of Technology, Turkey
From page
1169
To page
1175
Abstract
Approximately 80% of newborn in Turkey are put in neonatal incubators because of their problematic cases. Incubators used for treatment may adversely affect baby’s health seriously,if they adjusts or measures the parameters incorrectly. In this study,complications arisen because of inaccurate adjustment and measurement of incubator parameters were investigated. Current infant incubator analyzers were researched and the deficiencies were evaluated considering the standards and clinical metrology studies. As a result,findings were revealed as one of the current two incubator analyzers don’t ensure the standard and the other one doesn’t analyze some parameters which affect the baby’s health directly.
Keywords
Infant Incubator , Infant Incubator Analyzer , Neonatal
Journal title
Gazi University Journal Of Science
Journal title
Gazi University Journal Of Science
Record number
2600780