Title of article
Sensitivity Analysis of the Integral Quality Monitoring SystemD Using Monte Carlo Simulation
Author/Authors
Oderinde, Oluwaseyi M Department of Medical Physics - University of the Free State - Bloemfontein, South Africa , du Plessis, F. C. P Department of Medical Physics - University of the Free State - Bloemfontein, South Africa
Pages
12
From page
1
To page
12
Abstract
The Integral Quality Monitoring (IQM) System is a real-time beam output verifying system that validates the integrity and accuracy
of patient treatment plan (TP) data during radiation treatment. The purpose of this study was to evaluate the sensitivity of the IQM
to errors in segment using EGSnrc/BEAMnrc Monte Carlo (MC) codes. Sensitivity analysis (SA) techniques were applied to study
the significance of small alterations of field sizes (segments) on the IQM signal response. One hundred and eighty multileaf segments
were analyzed with methods that include scatter plots (SP), brute force, variance-based (VAR), and standard regression coefficient
SA. The segments were altered randomly within ±1, ±2, and ±3 mm leaf steps for 10 MV photon beams. SP analysis gradient and
VAR maximum index are 1.045 and 0.556 for the smallest segment while the largest segment has the value of 0.018 and 0.504,
respectively. The brute force and standard regression displayed maximum sensitivity indices around the unaltered segments. These
tests conclusively indicated that the IQM was more sensitive to alterations of small segments compared to larger segments. This is
important since small segment variation will cause a higher dose output variation that should be picked up during online beam
monitoring.