• Title of article

    Worse health-related quality of life and hip function in female patients with elevated chromium levels

  • Author/Authors

    HUSSEY, Daniel K Harris Orthopaedic Laboratory - Massachusetts General Hospital , MADANAT, Rami Harris Orthopaedic Laboratory - Massachusetts General Hospital , DONAHUE, Gabrielle S Harris Orthopaedic Laboratory - Massachusetts General Hospital , ROLFSON, Ola Harris Orthopaedic Laboratory - Massachusetts General Hospital , MURATOGLU, Orhun K Harris Orthopaedic Laboratory - Massachusetts General Hospital , MALCHAU, Henrik Harris Orthopaedic Laboratory - Massachusetts General Hospital

  • Pages
    7
  • From page
    485
  • To page
    491
  • Abstract
    Background and purpose — Blood metal ion levels can be an indi- cator for detecting implant failure in metal-on-metal (MoM) hip arthroplasties. Little is known about the effect of bilateral MoM implants on metal ion levels and patient-reported outcomes. We compared unilateral patients and bilateral patients with either an ASR hip resurfacing (HR) or an ASR XL total hip replacement (THR) and investigated whether cobalt or chromium was associ- ated with a broad spectrum of patient outcomes. Patients and methods — From a registry of 1,328 patients enrolled in a multicenter prospective follow-up of the ASR Hip System, which was recalled in 2010, we analyzed data from 659 patients (311 HR, 348 THR) who met our inclusion criteria. Cobalt and chromium blood metal ion levels were measured and a 21-item patient-reported outcome measures (PROMs) question- naire was used mean 6 years after index surgery. Results — Using a minimal threshold of > 7 ppb, elevated chro- mium ion levels were found to be associated with worse health- related quality of life (HRQoL) (p < 0.05) and hip function (p < 0.05) in women. These associations were not observed in men. Patients with a unilateral ASR HR had lower levels of cobalt ions than bilateral ASR HR patients (p < 0.001) but similar levels of chromium ions (p = 0.09). Unilateral ASR XL THR patients had lower chromium and cobalt ion levels (p < 0.005) than bilateral ASR XL THR patients. Interpretation — Chromium ion levels of > 7 ppb were associ- ated with reduced functional outcomes in female MoM patients.
  • Keywords
    female patients , elevated chromium levels , health-related quality of life and hip function
  • Journal title
    Acta Orthopaedica
  • Serial Year
    2016
  • Record number

    2618303