• Title of article

    Validating dose rate calibration of radiotherapy photon beams through IAEA/WHO postal audit dosimetry service

  • Author/Authors

    Jangda, Abdul Qadir Aga Khan University Hospital - Radiation Oncology Department (IBN ZUHR), Pakistan , Hussein, Sherali Aga Khan University Hospital - Radiation Oncology Department (IBN ZUHR), Pakistan

  • From page
    490
  • To page
    493
  • Abstract
    In external beam radiation therapy (EBRT), the quality assurance (QA) of the radiation beam is crucial to the accurate delivery of the prescribed dose to the patient. One of the dosimetric parameters that require monitoring is the beam output, specified as the dose rate on the central axis under reference conditions. The aim of this project was to validate dose rate calibration of megavoltage photon beams using the International Atomic Energy Agency (IAEA)/World HealthOrganisation (WHO) postal audit dosimetry service. Three photon beams were audited: a 6 MV beam from the lowenergy linac and 6 and 18 MV beams from a dual high-energy linac. The agreement between our stated doses and the IAEA results was within 1% for the two 6 MV beams and within 2% for the 18 MV beam. The IAEA/WHO postal audit dosimetry service provides an independent verification of dose rate calibration protocol by an international facility.
  • Keywords
    Radiation , ERBT , Photon beams.
  • Journal title
    Journal of the Pakistan Medical Association (Centre) JPMA
  • Journal title
    Journal of the Pakistan Medical Association (Centre) JPMA
  • Record number

    2652213