Title of article
An On-Line Preconcentration System Based on Graphitic Carbon Nitride Coated Membrane Filter for the Fast Determination of Trace Levels of Lead in Water Samples
Author/Authors
Zolfonoun, Ehsan Nuclear Fuel Cycle Research School - Nuclear Science & Technology Research Institute, Tehran , Yousefi, Reza Nuclear Fuel Cycle Research School - Nuclear Science & Technology Research Institute, Tehran
Pages
7
From page
30
To page
36
Abstract
A ligandless on-line solid phase extraction method was developed for the preconcentration of lead prior to
quantitation by inductively coupled plasma-optical emission spectrometry. In this method, the sample solution was
passed through a syringe membrane filter coated with graphitic carbon nitride and Pb(II) ions were directly adsorbed
on the surface of g-C3N4 particles. The adsorbed Pb(II) ions were subsequently eluted from the membrane and
transferred directly into the ICP-OES nebulizer with nitric acid solution. Under the optimized conditions and
preconcentration of 25 mL of sample, the enhancement factor of 110 and the detection limit of 0.12 μg L-1 were
obtained. The proposed procedure was applied for the preconcentration and determination of lead in environmental
water samples.
Keywords
On-Line Preconcentration , Solid Phase Extraction , Graphitic Carbon Nitride , Lead , ICP-OES
Journal title
Iranian Journal of Analytical Chemistry
Serial Year
2020
Record number
2703515
Link To Document