    Energy Window Selection for Bremsstrahlung 90Y SPECT-CT Imaging: A Phantom Study

  • Author/Authors

    Curto ، Denise Physics Department - University of Trieste , Bonutti ، Faustino Medical Physics Department - Academic Hospital of Udine , Bouzekraoui ، Youssef Laboratory of Sciences and Health Technologies - High Institute of Health Sciences - Hassan First University of Settat , Bentayeb ، Farida Department of Physics - Faculty of Science - Mohammed V University , Asmi ، Hicham Department of Physics - Faculty of Science - Mohammed V University

  • From page
    43
  • To page
    48
  • Abstract
    Introduction: In Yttrium-90 SPECT imaging, the energy window and collimator used during projection acquisition can significantly affect image quality. In this work, we used a new and independent method to verify previous results, which suggest suitable energy around 130 keV. Material and Methods: We used Siemens Symbia SPECT-CT system fitted with High Energy General Purpose (HEGP), Medium Energy General Purpose (MEGP), and Low Energy High Resolution (LEHR) to acquire data from NEMA IEC PET Body Phantom filled with 90Y chloride. ISO-counting curve is a new method analysed in this study to evaluate the adequate parameters for 90Y SPECT imaging. Results: HEGP collimator was the most suitable for acquisitions of 90Y bremsstrahlung radiation from the point of view of the correct volume reproduction. ISO-counting analyses have shown that for the bigger phantom spheres, the optimum acquisition energy is centered on 130 keV. Conclusion: The ISO-counting curve method is consistent to previous studies’ results and can help to improve image quality.
  • Keywords
    Bremsstrahlung , Yttrium , 90 , NEMA IEC PET Body , Phantom , ISO , Counting Curves
  • Journal title
    Iranian Journal of Medical Physics (IJMP)
  • Journal title
    Iranian Journal of Medical Physics (IJMP)
  • Record number

    2711889