Title of article
Probability Assessment of Photoneutron Production of Varian-Clinac iX Linear Accelerator Using Monte Carlo Simulation
Author/Authors
Zare Ganjaroodi ، Saeed Department of Physics and Energy Engineering - School of Sciences - Amirkabir University of Technology , Fathalizadeh ، Mohammadreza Faculty of Engineering - Islamic Azad University, Tehran Science and Research Branch , Zarifi ، Ehsan Nuclear Science and Technology Research Institute (NSTRI)
From page
37
To page
44
Abstract
Overdose in the patient’s body disrupts the treatment process and may have devastating effects. On the other hand, if the particle is a neutron, this effect would be multiplied. Because neutrons in the energy range of about 0.1 to 2 MeV produced in medical linear accelerators have a Quality Factor (QA) of 20, which creates a high equivalent dose in the tissue. In this paper, the photoneutron production probability of components of the 18 MV Varian-Clinac ix linear accelerator has been performed using the Monte Carlo simulation. The contribution of each gantry component of the accelerator and phantom in neutron production per gray of photons was calculated. Results showed that the largest ratio in the production of photoneutrons belongs to the primary collimator per photon gray per square centimetre. Also, in the target, which is the first source of photon neutron production, the flux of thermal neutrons is calculated at zero.
Keywords
Neutron , Varian , Clinac ix , Production probability , Monte Carlo simulation
Journal title
International Journal of Reliability, Risk and Safety: Theory and Application
Journal title
International Journal of Reliability, Risk and Safety: Theory and Application
Record number
2781272
Link To Document