Title of article
Gas Chromatographic Determination of 1,4-Dioxane at Low Parts-Per-Million Levels in Glycols
Author/Authors
Pundlik، M.D. نويسنده , , Sitharaman، B. نويسنده , , Kaur، Inderjit نويسنده ,
Issue Information
ماهنامه با شماره پیاپی سال 2001
Pages
-72
From page
73
To page
0
Abstract
1,4-Dioxane is a flammable liquid and tends to form explosive peroxides. Its formation in glycols (low parts-per-million levels), which are used as dehumidifying agents in refineries, may take place by condensation. 1,4-Dioxane thus formed gets distilled over with benzene in the refinery process. Therefore, it is necessary to identify and determine the levels of 1,4-dioxane in glycols as well as benzene. Gas chromatography (GC) is probably the best technique for this purpose. GC analysis may be carried out using a flame ionization detector. Results show that 1,4-dioxane can be comfortably determined down to 2 ppm in glycols and benzene.
Keywords
Fullerenes , Organic compounds , Chemical synthesis , Infrared spectroscopy , Electronic paramagnetic resonance (EPR)
Journal title
JOURNAL OF CHROMATOGRAPHIC SCIENCE
Serial Year
2001
Journal title
JOURNAL OF CHROMATOGRAPHIC SCIENCE
Record number
37749
Link To Document