Title of article
Accurate assessment of patellar tracking using fiducial and intensity-based fluoroscopic techniques
Author/Authors
T. S. Y. Tang، نويسنده , , N. J. MacIntyre، نويسنده , , H. S. Gill، نويسنده , , R. A. Fellows، نويسنده , , N. A. Hill، نويسنده , , D. R. Wilson، نويسنده , , R. E. Ellis، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2004
Pages
9
From page
343
To page
351
Abstract
Accuracies of a point-based and an intensity-based fluoroscopic methods of assessing patella tracking were determined by comparing the pattern of patellar motion with respect to orientation (flexion, internal rotation, and lateral tilt) and translation (lateral, proximal, and anterior) with the pattern of patellar motion measured using Roentgen stereophotogrammetric analysis in three cadaver knee specimens. Each pose in the patellar motion could be obtained from single as well as multiple calibrated fluoroscopic images. The errors using the intensity-based method were slightly higher than those of the point-based method, but they appear to be sufficiently low to detect clinically significant differences in patellar kinematics.
Keywords
2D/3D registration , Patellar kinematics
Journal title
Medical Image Analysis
Serial Year
2004
Journal title
Medical Image Analysis
Record number
449842
Link To Document