Title of article
Steady flow dynamics of prosthetic aortic heart valves: a comparative evaluation with PIV techniques
Author/Authors
W. L. Lim، نويسنده , , Y. T. Chew، نويسنده , , T. C. Chew، نويسنده , , H. T. Low، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1998
Pages
11
From page
411
To page
421
Abstract
Particle Image Velocimetry (PIV), capable of providing full-field measurement of velocities and flow stresses, has become an invaluable tool in studying flow behaviour in prosthetic heart valves. This method was used to evaluate the performances of four prosthetic heart valves; a porcine bioprostheses, a caged ball valve, and two single leaflet tilting disc valves with different opening angles. Flow visualization techniques, combined with velocity vector fields and Reynolds stresses mappings in the aortic root obtained from PIV, and pressure measurements were used to give an overall picture of the flow field of the prosthetic heart valves under steady flow conditions. The porcine bioprostheses exhibited the highest pressure loss and Reynolds stresses of all the valves tested. This was mainly due to the reduction in orifice area caused by the valve mounting ring and the valve stents. For the tilting disc valves, a larger opening angle resulted in a smoother flow profile, and thus lower Reynolds stresses and pressure drops. The St. Vincent valve exhibited the lowest pressure drop and Reynolds stresses.
Keywords
Prosthetic heart valves , Comparative analysis , Reynolds stresses , Particle image velocimetry
Journal title
Journal of Biomechanics
Serial Year
1998
Journal title
Journal of Biomechanics
Record number
450690