  • Abstract
    We present a newly developed torsional stiffness measurement device with the potential to quantitatively assess the in vivo torsional stiffness of bone regenerate during distraction osteogenesis. We describe the form and function of this device and its application in a model of regenerate consolidation. The device was able to produce data to assess stiffness of the regenerate with an accuracy between ±3 and ±9% for material stiffness ranging between 0.1 and 2.4 Nm/° and with a precision of ±3.6%. This method provides advantages over similar methods of bone fracture healing assessment with guaranteed maintenance of bone axis, minimized risk of bone misalignment during the bone healing process and a close relation to the functional loading pattern in torsion of bones such as tibia and femora.
  • Keywords
    Bone torsional sti!ness , Quantitative bone healing assessment , External "xation , Bone healing , Distraction osteogenesis