• Title of article

    Design and validation of a surrogate humerus for biomechanical testing

  • Author/Authors

    A. P. Elfick، نويسنده , , G. Bedi، نويسنده , , A. Port، نويسنده , , A. Unsworth، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2002
  • Pages
    4
  • From page
    533
  • To page
    536
  • Abstract
    At present biomechanical testing of fracture plating strategies is conducted using animal or cadaveric whole bone models. This may introduce experimental error into these studies. This communication summarises the design and validation of a novel bone and fibre-reinforced plastic construct conceived to minimise intra-experimental error. A tubular surrogate humerus was produced with dimension and strength matched to that of the human humerus. Bone inserts placed into the wall of the tube allow for the fixation of the plates with bone screws. Three-point bending tests of the flexural rigidity of the surrogate humerus (EI=100.1 (SD 6.0) Nm2) showed it to be comparable to the human humerus. Further, pull-out tests of the screws showed that the bone slots adequately mimicked the whole bone scenario. This testing construct will be used for a comparative study of humeral plating techniques.
  • Keywords
    Cortical screw , Pull-out test , Biomechanical testing , Proxy humerus
  • Journal title
    Journal of Biomechanics
  • Serial Year
    2002
  • Journal title
    Journal of Biomechanics
  • Record number

    451288