• Title of article

    Analysis of the transient expansion behavior and design optimization of coronary stents by finite element method

  • Author/Authors

    Wei-Qiang Wang، نويسنده , , Dong-Ke Liang، نويسنده , , Da-Zhi Yang، نويسنده , , Min Qi، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2006
  • Pages
    12
  • From page
    21
  • To page
    32
  • Abstract
    The percutaneous transluminal coronary angioplasty (PTCA) assisted with stenting technique has become a primary therapy to coronary heart disease. In practice, the structure conditions of both ends of stent/balloon system influence a stentʹs instantaneous expansion behavior. The transitory nonuniform expansion, the so-called dogboning, of stent/balloon system is one of the main reasons to induce the acute vascular injury at the two edges of a stent. This kind of vascular injury has a close relationship with the in-stent restenosis. In the present paper, the finite element method (FEM) was applied to simulate the transient expansion process of stent/balloon system with different stent structure and balloon length under the internal pressure. And two types of stent and six collocations of stent and balloon were modeled. Modeling results showed that the dogboning phenomenon can be eliminated by improving geometry of a stent or/and varying the length of balloon over stent. The above modeled results were further confirmed by following in situ observation.
  • Keywords
    Stent , Finite element method , Design optimization , Foreshortening , Dogboning
  • Journal title
    Journal of Biomechanics
  • Serial Year
    2006
  • Journal title
    Journal of Biomechanics
  • Record number