Title of article
Transient elastohydrodynamic lubrication analysis of metal-on-metal hip implant under simulated walking conditions
Author/Authors
F. Liu، نويسنده , , Z.M. Jin، نويسنده , , F. Hirt، نويسنده , , C. Rieker، نويسنده , , P. Roberts، نويسنده , , P. Grigoris، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2006
Pages
10
From page
905
To page
914
Abstract
The transient elastohydrodynamic lubrication (EHL) analysis was performed in this study for a typical metal-on-metal bearing employing a polyethylene backing underneath a metallic cup inlay under dynamic operating conditions of load and speed representative of normal walking. A ball-in-socket configuration was adopted to represent the articulation between the femoral head and the acetabular cup. The governing Reynolds and elasticity equations were solved simultaneously by using both finite difference and finite element methods. The predicted transient film thickness from the present study was compared with the estimation based on the quasi-static analysis. It was found that the polyethylene backing employed in the typical metal-on-metal hip bearing, combined with dynamic squeeze-film action, significantly improved the transient lubricant film thickness under cyclic walking and consequently a fluid film lubrication regime was possible for smooth bearing surfaces with an average roughness less than 0.005 μm
Keywords
UHMWPE backing , Transient elastohydrodynamic lubrication , Metal-on-metal hip prosthesis
Journal title
Journal of Biomechanics
Serial Year
2006
Journal title
Journal of Biomechanics
Record number
453467
Link To Document