• Title of article

    Prevalence and clinical impact of stent fractures after femoropopliteal stenting Original Research Article

  • Author/Authors

    Dierk Scheinert، نويسنده , , Susanne Scheinert، نويسنده , , Jacqueline Sax، نويسنده , , Christopher Piorkowski، نويسنده , , Sven Br?unlich، نويسنده , , Matthias Ulrich، نويسنده , , Giancarlo Biamino، نويسنده , , Andrej Schmidt، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2005
  • Pages
    4
  • From page
    312
  • To page
    315
  • Abstract
    Objectives The aim of this study was to investigate the occurrence and the clinical impact of stent fractures after femoropopliteal stenting. Background The development of femoral stent fractures has recently been described; however, there are no data about the frequency and the clinical relevance. Methods A systematic X-ray screening for stent fractures was performed in 93 patients. In total, 121 legs treated by implantation of self-expanding nitinol stents were investigated after a mean follow-up time of 10.7 months. The mean length of the stented segment was 15.7 cm. Results Overall, stent fractures were detected in 45 of 121 treated legs (37.2%). In a stent-based analysis, 64 of 261 stents (24.5%) showed fractures, which were classified as minor (single strut fracture) in 31 cases (48.4%), moderate (fracture of >1 strut) in 17 cases (26.6%), and severe (complete separation of stent segments) in 16 cases (25.0%). Fracture rates were 13.2% for stented length ≤8 cm, 42.4% for stented length >8 to 16 cm, and 52.0% for stented length >16 cm. In 21 cases (32.8%) there was a restenosis of >50% diameter reduction at the site of stent fracture. In 22 cases (34.4%) with stent fracture there was a total stent reocclusion. According to Kaplan-Meier estimates, the primary patency rate at 12 months was significantly lower for patients with stent fractures (41.1% vs. 84.3%, p < 0.0001). Conclusions There is a considerable risk of stent fractures after long segment femoral artery stenting, which is associated with a higher in-stent restenosis and reocclusion rate.
  • Keywords
    SFA , superficial femoral artery
  • Journal title
    JACC (Journal of the American College of Cardiology)
  • Serial Year
    2005
  • Journal title
    JACC (Journal of the American College of Cardiology)
  • Record number

    459688