• Title of article

    Interventional Atrial Septal Defect Closure Using a Totally Bioresorbable Occluder Matrix: Development and Preclinical Evaluation of the BioSTAR Device Original Research Article

  • Author/Authors

    Christian Jux، نويسنده , , Harald Bertram، نويسنده , , Peter Wohlsein، نويسنده , , Michael Bruegmann، نويسنده , , Thomas Paul، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2006
  • Pages
    9
  • From page
    161
  • To page
    169
  • Abstract
    Objectives We sought to test the hypothesis that interventional atrial septal defect (ASD) closure can be performed safely and effectively using a bioresorbable occluder matrix. Background The ideal septal occluder scaffold should promote the healthiest and most complete healing response while eventually facilitating the full resorption of the material and leaving “native” tissue behind, thus minimizing the potential for future complications from chronic foreign body and maintaining the possibility for later unobstructed transseptal access to the left atrium. Methods The STARFlex occluders (NMT Medical Inc., Boston, Massachusetts) were modified by substituting the conventional polyester fabric for a bioengineered, acellular type-I collagen matrix derived from porcine submucosa with a heparin-coated surface (BioSTAR occluder, NMT Medical Inc.). Comparative transcatheter closure of ASDs was performed in young sheep (n = 36). Gross pathology and histopathology were obtained after follow-up periods ranging from 7 days to 2 years. Results The STARFlex (control) devices were encapsulated time-dependently by ingrown fibrous tissue. Histology showed a mild but chronically persisting foreign body reaction. By contrast, BioSTAR devices exhibited a mild-to-moderate transient cellular immune response. Heparin coating of the BioSTAR surface improved the biocompatibility of the device by reducing surface thrombogencity. A remodeling process of the collagen scaffold, starting after 30 days in vivo, resulted in the full replacement of the matrix by host tissue after 2 years of follow-up. Conclusions The BioSTAR device is the first septal occluder with a totally bioresorbable matrix that is fully replaced by host tissue during the healing process. The promising results of this study support testing of the BioSTAR device in clinical trials.
  • Keywords
    ASD , ice , Icl , Atrial septal defect , intracardiac echocardiography , intestinal collagen layer
  • Journal title
    JACC (Journal of the American College of Cardiology)
  • Serial Year
    2006
  • Journal title
    JACC (Journal of the American College of Cardiology)
  • Record number

    471863