• Title of article

    Automation of toxicological screenings on a Hewlett Packard Chemstation GC-MS system

  • Author/Authors

    Tom Fiers، نويسنده , , Viviane Maes، نويسنده , , Claude Sevens، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1996
  • Pages
    5
  • From page
    357
  • To page
    361
  • Abstract
    Objectives: To optimize the interpretation of GC-MS toxicological screenings (i.e., to facilitate ion specific queries, create custom reports specifically adapted to each confirmation procedure, and eliminate redundant and/or inaccurate data on library search reports). Design and Methods: The MS Chemstation software of the Hewlett Packard 5972 is constructed in a modular way. We made extensive modifications to two modules, the data analysis and the report modules, using the built-in MS Chemstation macro language. Results: Ion specific queries were automated for over 60 commonly encountered analytes. Custom reports were created for the confirmation of positive drugs-of-abuse immunoassay results. With the incorporation of decision support rules into the data processing and the reporting phases, we obtained sensitive, accurate, and concise reports. Conclusions: The MS Chemstation software can be tailored to the needs of each individual application. The incorporation of a rule-based decision support system enhances the quality of the GC-MS toxicological screenings and results in faster, easier, and more reliable processing.
  • Keywords
    Expert system , Automation , GC-MS , software design , Decision Analysis , Toxicology , Hewlett Packard
  • Journal title
    Clinical Biochemistry
  • Serial Year
    1996
  • Journal title
    Clinical Biochemistry
  • Record number

    481605