• Title of article

    Assay of neonatal TSH from blood spots using the Architect i2000, an immunoassay analyzer

  • Author/Authors

    Joyce LePage، نويسنده , , Sherry Peters، نويسنده , , Denis C. Lehotay، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2004
  • Pages
    3
  • From page
    415
  • To page
    417
  • Abstract
    Objectives: To assess the Architect i2000 for measuring TSH from neonatal blood spots. Methods: Blood collected on filter paper was eluted and assayed on the Architect. Results: Recovery of TSH was close to 100%. Within-run and between-run precisions were 4.2–8.2% and 14.5–18.1%. Correlation with the Wallac Delphia method was excellent (r = 0.96). Linearity, sensitivity, and slope were 0–400 mU/l, 0.15 mU/l, and 0.86, respectively. The results obtained by this method on the Architect allowed us to differentiate normals from affected patients with the same reliability as the previous method. Conclusions: Measurement of TSH on the Architect is an excellent method for detecting congenital hypothyroidism.
  • Keywords
    TSH , Architect i2000 , Radioimmunoassays
  • Journal title
    Clinical Biochemistry
  • Serial Year
    2004
  • Journal title
    Clinical Biochemistry
  • Record number

    482534