Title of article
Inflammatory cytokine removal by an activated carbon device in a flowing system
Author/Authors
Susan R. Sandeman، نويسنده , , Carol A. Howell، نويسنده , , Sergey V. Mikhalovsky، نويسنده , , Gary J. Phillips، نويسنده , , Andrew W. Lloyd، نويسنده , , J. Graham Davies، نويسنده , , Stephen R. Tennison، نويسنده , , Anthony P. Rawlinson، نويسنده , , Oleksaudr P. Kozynchenko، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2008
Pages
7
From page
1638
To page
1644
Abstract
A prototype in-line filtration/adsorption device has been developed using novel synthetic pyrolysed carbon monoliths with controlled mesoporous domains of 2–50 nm. Porosity was characterized by SEM and porosimetry. Removal of inflammatory cytokines TNF, IL-6, IL-1β and IL-8 was assessed by filtering cytokine spiked human plasma through the walls of the carbon modules under pressure. The effect of carbon filtration on plasma clotting response and total plasma protein concentration was also assessed. Significant removal of the cytokines IL-6, IL-1β and IL-8 was observed. Initially marked TNF removal diminished over time. The coagulation studies indicated that the carbon device does not exacerbate the propensity of blood plasma to clot. The total plasma protein concentration remained constant. The device offers a broader approach to the treatment of systemic inflammatory response syndrome (SIRS) by the removal of inflammatory mediators central to its progression.
Keywords
inflammation , cytokine , Adsorption , Clotting
Journal title
Biomaterials
Serial Year
2008
Journal title
Biomaterials
Record number
482967
Link To Document