• Title of article

    Novel comparison of capillary electrophoresis versus immunoassay in the measurement of total percent carbohydrate deficient transferrin

  • Author/Authors

    Kathy A. Strauss، نويسنده , , Jessica OʹKane، نويسنده , , Janine Denis Cook، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2006
  • Pages
    7
  • From page
    826
  • To page
    832
  • Abstract
    Objectives: Novel comparison of CDT isoforms as determined by CE with an FDA-approved immunoassay kit. Design and methods: Subjects (n = 51) were categorized by drinking status based on AUDIT questionnaire responses. CDT isoform analyses by CE were compared to a commercially available, FDA-approved immunoassay. The analytical specificity of the immunoassay kit was assessed by analysis with IEF. Results: Because of the poor correlation between % CDT as measured by CE and the TIA immunoassay and between subject-reported drinking levels and results from the TIA assay, extraction column eluants from the immunoassays were analyzed by IEF for analytical specificity. % CDT by TIA included some trisialo-Tf, a non-CDT fraction, in the % CDT determination. Conclusions: Total % CDT by CE, which separates all isoforms is more analytically specific than immunoassays because it does not include trisialo-Tf in the CDT calculation.
  • Keywords
    alcohol , Capillary electrophoresis (CE) , Carbohydrate deficient transferrin (CDT)
  • Journal title
    Clinical Biochemistry
  • Serial Year
    2006
  • Journal title
    Clinical Biochemistry
  • Record number

    484797