Title of article
Extraction-flow injection spectrofluorimetric measurement of dissolved oxygen in environmental waters using 2-thionaphthol
Author/Authors
Sakai، Tadao نويسنده , , Takio، Hisashi نويسنده , , Teshima، Norio نويسنده , , Nishikawa، Harumitsu نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2001
Pages
-116
From page
117
To page
0
Abstract
A simple and highly sensitive flow injection analysis for the spectrofluorimetric determination of dissolved oxygen (DO) in environmental waters was developed. The decrease in fluorescence, which is based on the reaction of 2-thionaphthol with iodine liberated in Winklerʹs method, was utilized as the DO detection system. The calibration graph was linear over the range 2?10-6-1.2?10-5mol/l iodine. The relative standard deviation (n=8) was below 1% for the 6?10-6mol/l iodine determination and the detection limit (S/N=3) was 4.9? 10-7mol/l. The sample frequency was 18h-1.
Keywords
Beryllium determination , Aurintricarboxylic acid , solid-phase spectrophotometry , water analysis
Journal title
Analytica Chimica Acta
Serial Year
2001
Journal title
Analytica Chimica Acta
Record number
49462
Link To Document