Title of article
Antigen detection using microelectrode array microchips
Author/Authors
WANG، Wei نويسنده , , Dill، Kilian نويسنده , , Montgomery، Donald D. نويسنده , , Tsai، Julie C. نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2001
Pages
-68
From page
69
To page
0
Abstract
Procedures and results are described for multiplexed immunochemical assays using semiconductor microchips. The microchips used here are miniaturized arrays of individually addressable microelectrodes controlled by active CMOS circuitry. Electrode densities exceed 1000percm2. The array chips are coated with a porous reaction layer material to provide a `biofriendlyʹ milieu overlaying the electrode array. Biotin is linked covalently to regions within the porous reaction layer proximate to selected microelectrodes. Covalent linkage is accomplished using reagents that are generated in situ by the microelectrodes. The covalent linkage of biotin within the porous reaction layer allowed traditional streptavidin (SA)-based immunoassay formats to be used on the biochips.
Keywords
Glucuronides , ELISA , Integrated HPLC-ELISA , Immunoassay , phenolic compounds
Journal title
Analytica Chimica Acta
Serial Year
2001
Journal title
Analytica Chimica Acta
Record number
49657