• Title of article

    Flow injection preconcentration by chemofiltration and spectrophotometric determination of Gd

  • Author/Authors

    Raba، Julio نويسنده , , Vito، Irma E. De نويسنده , , Olsina، Roberto A. نويسنده , , Masi، Adriana N. نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2004
  • Pages
    -10
  • From page
    11
  • To page
    0
  • Abstract
    A new on-line Gd preconcentration and determination system associated to flow injection (FI) method was developed. 2,2ʹ-(1,8-dihydroxy-3,6-disulfonaphthylene2,7-bisazo) bisbenzenearsonic acid (Arsenazo III) was used as a complexing agent at pH 2.5. A reactor containing the polyamide membrane was used for the retention of the Gd complex by chemofiltration. The complex was then removed from the reactor with buffer solution pH 9. The variables affecting the combined online preconcentration–absorptiometric method have been evaluated and optimised. The coupling of the on-line preconcentration and spectrophotometric flow through detection led to a detection limit of 15 (mu)g l^-1 for a preconcentration time of 5 min at 2 ml/min. The method was successfully applied to biological samples.
  • Keywords
    Absorptiometric method , Preconcentration and determination , Gadolinium in urine sample , Flow Injection
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    2004
  • Journal title
    Analytica Chimica Acta
  • Record number

    49861