    Performance of a new laser fluorescence device for the detection of occlusal caries in vitro

  • Author/Authors

    A. Lussi، نويسنده , , E. Hellwig، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2006
  • Pages
    5
  • From page
    467
  • To page
    471
  • Abstract
    The new device DIAGNOdent pen based on red laser light induced fluorescence was introduced for the detection of approximal and occlusal caries. The aim of this study was to test its performance on occlusal surfaces. The new device comes with two different sapphire fibre tips: a cylindrical tip and a conical tip. The two new sapphire fibre tips were used and compared with the tip currently available with DIAGNOdent (DD). Methods The teeth were selected from a pool of extracted permanent human molars, which were stored frozen at −20 °C, until use. Prior to being measured the teeth were defrosted and cleaned. One hundred and nineteen teeth were selected and measured with the old tip and with the two new tips of the new device by two independent assessments. The teeth were histologically prepared and assessed for caries extension. Results Specificity values for D1, D2 and D3 ranged between 0.69 and 0.89, sensitivity between 0.78 and 0.96. There were no statistically significant differences obtained between the two tips of the new and the one tip of the old device (p > 0.05). Intra-examiner reliability with kappa values of >0.83 was high. Conclusions In this study, the new laser fluorescence device performed on occlusal surfaces as well as the available device.
  • Keywords
    Caries detectionOcclusal cariesLaser fluorescenceDIAGNOdentDIAGNOdent pen
  • Journal title
    Journal of Dentistry
  • Serial Year
    2006
  • Journal title
    Journal of Dentistry
  • Record number

    507521