• Title of article

    Sheathless Capillary Electrophoresis/Electrospray Mass Spectrometry Using a Carbon-Coated Tapered Fused-Silica Capillary with a Beveled Edge

  • Author/Authors

    Chang، Yan Zin نويسنده , , Chen، Yet Ran نويسنده , , Her، Guor Rong نويسنده ,

  • Issue Information
    دوهفته نامه با شماره پیاپی سال 2001
  • Pages
    -5082
  • From page
    5083
  • To page
    0
  • Abstract
    A tapered capillary tip containing a beveled edge was developed for use in sheathless capillary electrophoresis/electrospray mass spectrometry (CE/ESI-MS). The optimal flow rate of a 75-mu m-i.d., 90-mu m-o.d. beveled tapered capillary tip was similar to a conventional flat tapered tip with a 25-mu m orifice. Using a mixture of coptisine, berberine, and palmatine chloride, the sheathless CE/ESI-MS sensitivity of a beveled 75 mu m tapered tip capillary was found to be similar to a 25 mu m flat tip. Although both tips offer similar CE/ESI-MS sensitivity, the beveled tapered capillary tip is more rugged and durable than a conventional 25-mu m tapered capillary because of the larger outside diameter and inside diameter. To make electrical contact, the capillary tip was smeared with paint marker followed by the application of a carbon coating using a graphite pencil. Using this refined carbon-coating procedure, the capillary tip can be operated with aprotic solvents.
  • Keywords
    High-performance liquid chromatography-mass spectrometry (LC-MS) , Solid-phase extraction (SPE) , Trifolium pratense , red clover , Isoflavone glycoside 6"-O-malonates , Isoflavone glycoside 6"-O-acetates
  • Journal title
    Analytical Chemistry
  • Serial Year
    2001
  • Journal title
    Analytical Chemistry
  • Record number