• Title of article

    Clinical predictors of defibrillation energy requirements in patients treated with a nonthoracotomy defibrillator system

  • Author/Authors

    S. Adam Strickberger، نويسنده , , Sheldon L. Brownstein، نويسنده , , Bruce L. Wilkoff، نويسنده , , Alexander J. Zinner، نويسنده , , ResQ Investigators، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1996
  • Pages
    4
  • From page
    257
  • To page
    260
  • Abstract
    Many factors can influence defibrillation energy requirements (DER) in patients with a nonthoracotomy defibrillator. No large studies, however, have correlated clinical characteristics with the DER. In this study, 124 patients underwent the same DER protocol with the identical biphasic waveform, nonthoracotomy lead system, and lead configuration. These patients were 63 ± 12 years old (mean ± SD); 99 were men; the ejection fraction was 0.32 ± 0.13 and 36 were taking an antiarrhythmic medication. New York Heart Association congestive heart failure class I was present in 28, class II in 70, and class III in 26 patients. Male sex (454 ± 94 V vs 406 ± 91 V for female sex) was associated with a significantly higher DER (p = 0.02) and an increased risk of a DER>550 V (p = 0.047). No other clinical variable was associated with the DER or a DEr>550 V. In conclusion, women tend to have lower DERs than men.
  • Journal title
    American Heart Journal
  • Serial Year
    1996
  • Journal title
    American Heart Journal
  • Record number

    526831