Title of article
Prompt-gamma neutron activation facility for measuring body nitrogen in vivo in small animals
Author/Authors
I. E. Stamatelatos، نويسنده , , S. Yasumura، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1995
Pages
4
From page
269
To page
272
Abstract
We have designed a prompt-gamma neutron activation facility for in vivo measurements of total body N in small animals. The facility incorporates a 252C neutron source with an emission rate of 3.2 × 108 n s−1 (140 μg), and a heavy-water neutron collimator-moderator assembly, together providing an optimum thermal flux-to-dose ratio for a 0.25–0.5 kg animal. Using two NaI(Tl) detectors we can attain a fractional statistical error of measurement of 3.5% for a dose equivalent of approx 50 mSv, sufficiently low that serial measurements can be made on the same animal.
Journal title
Applied Radiation and Isotopes
Serial Year
1995
Journal title
Applied Radiation and Isotopes
Record number
539210
Link To Document