Title of article
A new method for automatic measurement of low-level radon in water
Author/Authors
Pall Theodorsson، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1996
Pages
5
From page
855
To page
859
Abstract
A new single phototube liquid scintillation counting system for automatic determination of 222Rn in ground water for earthquake prediction is described. About 40% of radon is transferred from 1.0 L of water with a stream of air into 25 ml of toluene in a sample vial. Delayed coincidence counting of 214Bi and its decay product, 214Po reduces the background to 0.02 cpm. The 214Po detection efficiency is above 90%. Water containing 20 mBq/L of radon can be measured with 10% accuracy in 3 h.
Journal title
Applied Radiation and Isotopes
Serial Year
1996
Journal title
Applied Radiation and Isotopes
Record number
539564
Link To Document