Title of article
Alanine EPR dosimeter response in proton therapy beams
Author/Authors
K. GALL، نويسنده , , M. Desrosiers، نويسنده , , D. Bensen، نويسنده , , C. Serago، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1996
Pages
3
From page
1197
To page
1199
Abstract
We report a series of measurements directed to assess the suitability of alanine as a mailable dosimeter for dosimetry quality assurance of proton radiation therapy beams. These measurements include dose-response of alanine at 140 MeV, and comparison of response vs energy with a parallel plate ionization chamber. All irradiations were made at the Harvard Cyclotron Laboratory, and the dosimeter were read at NIST. The results encourage us that alanine could be expected to serve as a mailable dosimeter with systematic error due to differential energy response no greater than 3% when doses of 25 Gy are used.
Journal title
Applied Radiation and Isotopes
Serial Year
1996
Journal title
Applied Radiation and Isotopes
Record number
539622
Link To Document