• Title of article

    ESR dosimeters based on metallodiphthalocyanines: Correlation between ESR and spectrophotometric data

  • Author/Authors

    E. A. Lukʹyanets، نويسنده , , E. V. Pykhtina، نويسنده , , L. A. Ulanova، نويسنده , , M. A. Kovaleva، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1996
  • Pages
    6
  • From page
    1541
  • To page
    1546
  • Abstract
    Ionizing radiation (IR) induced redox processes in liquid dosimetric systems based on diphthalocyanine metal (DPcM) complexes solutions in organic solvents are accompanied by changes in their visible spectra and magnetic properties which can be measured by electron spin resonance (ESR) method at g congruent with 2 and spectrophotometry (SP) without any special treatment of the samples. The good correlation between ESR and SP data for different systems based on solutions of DPc complexes of Lu, Hf and Zr can be explained by the fact that both methods determine the concentration of DPcM radical forms. The better accuracy of ESR method makes it possible to measure lower doses than using SP procedure.
  • Journal title
    Applied Radiation and Isotopes
  • Serial Year
    1996
  • Journal title
    Applied Radiation and Isotopes
  • Record number

    539681