• Title of article

    HPGe detector efficiency calibration for extended cylinder and marinelli-beaker sources using the ESOLAN program

  • Author/Authors

    Tien-Ko Wang، نويسنده , , Wei-Yang Mar، نويسنده , , Tzung-Hua Ying، نويسنده , , Chia-Lian Tseng، نويسنده , , Chi-Hung Liao، نويسنده , , Mei-Ya Wang، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1997
  • Pages
    13
  • From page
    83
  • To page
    95
  • Abstract
    A personal computer based program ESOLAN has been developed for HPGe detector absolute-peak-efficiency calibration for various source-detector arrangements. The applicability of the previously developed program is limited to sources with diameter smaller than that of the detector active crystal. In this work, the ESOLAN program is modified to extend its application to sources with dimensions larger than the detector crystal, including extended cylinder and Marinelli-beaker sources.
  • Journal title
    Applied Radiation and Isotopes
  • Serial Year
    1997
  • Journal title
    Applied Radiation and Isotopes
  • Record number

    539721