• Title of article

    Mixing activity-calibrated and uncalibrated gamma-ray sources in efficiency calibration

  • Author/Authors

    R. Tramontano، نويسنده , , V. R. Vanin، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1999
  • Pages
    6
  • From page
    323
  • To page
    328
  • Abstract
    A method to calibrate the absolute efficiency of HPGe detectors in a large energy range using both activity-calibrated and uncalibrated multienergy gamma-ray radioactive sources is presented. The adopted efficiency calibration function var epsilon(E) has the simple form ln var epsilon(E)=a0+a1 ln(E/EB)+a2 ln2(E/EB). Different parameters a0 are simultaneously fitted to the activity-calibrated and uncalibrated sources experimental efficiency data by the least-squares method. As an application example, the calibration of a HPGe detector in the 500 to 3500 keV energy range using a 207Bi activity calibrated source and one uncalibrated 56Co source is described.
  • Keywords
    gamma radiation
  • Journal title
    Applied Radiation and Isotopes
  • Serial Year
    1999
  • Journal title
    Applied Radiation and Isotopes
  • Record number

    540459