    Measurement of 49V half-life

  • Author/Authors

    L. Rodriguez Barquero، نويسنده , , J. M. Los Arcos، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2000
  • Pages
    3
  • From page
    569
  • To page
    571
  • Abstract
    The accepted value for the 49V half-life is either 330±15 days, based on a few old measurements with large dispersion, or 338±5 days, from a more recent measurement. Methods of study have included proportional counting and NaI X-ray detection. A new direct measurement of the 49V half-life has been carried out for over 2.5 half lives using liquid scintillation counting, as an additional contribution due to the scarce measurements of this radionuclide. A half-life value of 328.6±3.3 days was determined on the basis of an overall uncertainty analysis of 0.99%.
  • Journal title
    Applied Radiation and Isotopes
  • Serial Year
    2000
  • Journal title
    Applied Radiation and Isotopes