Title of article
Accurate computation of coincidence summing corrections in low level gamma-ray spectrometry
Author/Authors
Octavian Sima، نويسنده , , Dirk Arnold، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2000
Pages
6
From page
51
To page
56
Abstract
The GESPECOR (Germanium Spectrometry Correction factors) software, previously developed for computing the self-attenuation and coincidence summing corrections, was applied to the computation of the coincidence summing correction factors for a well-type and two coaxial HPGe detectors. Cylindrical samples as well as Marinelli beaker samples were considered. The computed values are in good agreement with carefully measured values. A detailed study of the uncertainties assigned to the results was carried out. The analysis shows that the procedures used in GESPECOR are reliable and provide results with a well defined accuracy.
Keywords
Germanium detectors , Coincidence summing , Efficiency calibration , Low level gamma-ray spectrometry
Journal title
Applied Radiation and Isotopes
Serial Year
2000
Journal title
Applied Radiation and Isotopes
Record number
540724