• Title of article

    Computed tomography with monochromatic bremsstrahlung radiation

  • Author/Authors

    R. T. Lopes، نويسنده , , E. B. Costa، نويسنده , , E. F. O. De Jesus، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2000
  • Pages
    7
  • From page
    665
  • To page
    671
  • Abstract
    The use of filters with appropriate thickness of different elements produces a method to quasi-monochromatize X-ray beams. In the present study the filters were produced in a boric acid matrix, using the elements Mo, Sb, Cs, Ce and Gd. The thickness of the filters varied from 50 to 300 μm for a high voltage tube operating from 30 to 70 kVp. The parameters analyzed were the beam resolution, measured by its FWHM as a function of filter thickness, and the reduction of beam intensity. The results show a comparison between conventional tomographies obtained with different filters and application in the technique of differential tomography, with double energy for identification of elements such as Zr, Nd, Ba and Sm.
  • Keywords
    computed tomography , Bremsstrahlung radiation , Monoenergetic radiation
  • Journal title
    Applied Radiation and Isotopes
  • Serial Year
    2000
  • Journal title
    Applied Radiation and Isotopes
  • Record number

    540814