• Title of article

    Sensitive measurement of positron emitters eluted from HPLC

  • Author/Authors

    Makoto Takei، نويسنده , , Takayo Kida، نويسنده , , Kazutoshi Suzuki، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2001
  • Pages
    6
  • From page
    229
  • To page
    234
  • Abstract
    For sensitive analysis of the radioactive-metabolite in human PET, a radio-HPLC system coupled to a newly designed positron detector was constructed. The detector had the advantages of low noise level (1.7±1.0 cpm) and high sensitivity (32±1%) due to coincidence counting and large BGO crystals. Furthermore, the detector was easy to move, since a pair of the BGO housings coupled to photomultipliers was effectively arranged in parallel and a HPLC cell with different volume could be inserted between the BGO housing. This radio-HPLC system was useful for analyzing samples with low radioactivity. It was applied to the measurement of [11C]FLB457 in plasma, having high affinity and high selectivity with dopamine D2 receptors. Extremely low radioactivity of [11C]FLB457 (2500 dpm) could be analyzed by using the radio-HPLC system. The performance of this detector was compared with those of commercially available systems that had been used as sensitive detectors for HPLC.
  • Keywords
    Metabolite analysis , HPLC , Positron detector , Detection limit , sensitivity , PET
  • Journal title
    Applied Radiation and Isotopes
  • Serial Year
    2001
  • Journal title
    Applied Radiation and Isotopes
  • Record number

    541020